The Dropometer measured the surface tension of the hydrogel inkjet ink by pendant drop. The paper names it as the Dropometer pendant droplet analyzer from Droplet Lab, Canada.
Before surfactant, the prepolymerized ink measured 70 mN/m, which the authors judged not printable. After adding 2.4 µL of Triton X-100, it measured about 30 mN/m, which they judged printable.
The Dropometer did not measure viscosity; a separate A&D viscometer did. It also measured no contact angles, layer thickness or electrochemical response, which drive the paper's main conclusions.
The paper does not report drop volume, needle size, temperature, replicates, drop age or fitting model for the surface tension readings.
The surface tension reading confirmed when the ink reached a printable range, one of two rheology checks before printing. It supports ink formulation, not the pH switchable sensing result.