Engineering Functional Microcapsules for Controlled Cargo Delivery

This study develops fluorophilic boronic acid-stabilized hydrocarbon-fluorocarbon-water emulsions and uses pendant-drop PFH-water interfacial tension measurements to compare additive combinations that lead to stable, thermally reconfigurable droplets.

At-a-Glance Summary

Dropometer attribution in the paper

The study states that surface tensions were measured “using pendant drop method on the droplet lab tensiometer.”

How the surface-tension data were used in the study

These measurements were used to compare interface formulations. The thesis reports that the individual additives and the combined condition all lowered the interfacial tension, while lasting stabilization against coalescence was obtained only with the combined FBA-PVA system.

Paper Details

Title
Engineering Functional Microcapsules for Controlled Cargo Delivery (Chapter 3: Adaptive Stimuli-Triggered Response to Dynamic Environment: Thermally Reconfigurable Elastic Capsules)
Authors
Zhang Wu
Institution
Doctoral dissertation, Harvard University Graduate School of Arts and Sciences
Year
2025
Pages / Article
54–65
License
Other Posted Material (LAA)

What Was Measured

Primary surface / interfacial measurement

The study reports PFH-water interfacial tension from pendant-drop measurements with no additive, with 5 wt % FBA in PFH, with 10 wt % PVA in deionized water, and with the combined FBA-PVA condition. Figure 3.2 presents the pendant-drop images and the corresponding surface-tension values.

Supporting measurements

The interface interpretation is supported by transient wrinkle formation during drop retraction, DWR interfacial rheology of HFE-7500/water interfaces with different surfactants, and bright-field and fluorescent confocal microscopy of reconfigurable hexane-perfluorohexane elastic capsules at room temperature and around 0 ºC. The study also uses a glass capillary microfluidic device to fabricate the double-emulsion templates for the capsules.

Instruments Mentioned

Interfacial tension

droplet lab tensiometer

Interfacial rheology

HR 20 discovery hybrid rheometer by TA instruments

Interfacial rheology geometry

Double Wall Ring (DWR) setup

Droplet generation

glass capillary microfluidic device

Capsule morphology imaging

confocal microscopy

Role of the Dropometer

The study states that “the surface tensions of the interfaces are characterized using pendant drop method on the droplet lab tensiometer.” In practice, this was applied to PFH-water drop interfaces with and without FBA in the PFH phase and PVA in the aqueous phase, generating the interfacial-tension values reported in Figure 3.2.

Within the workflow, these Dropometer measurements were used to benchmark how each additive combination changed the PFH-water interface and, together with the author's coalescence observations, to indicate that similar or reduced interfacial tensions did not by themselves account for lasting fluorocarbon-water emulsion stabilization.

Bring This Measurement Into Your Lab

If you need to compare how additives change interfacial tension between oil and water, we can show you the pendant drop workflow on your liquids.

Method Snapshot

Method Snapshot Table

Measurement series Oil phase Aqueous phase Output reported Instruments Conditions Figure / notes
Baseline PFH-water interface PFH drop Deionized water (DIW) Surface tension: 71.6 mN/m droplet lab tensiometer PFH-water interface; pendant drop Fig. 3.2b
FBA in PFH PFH + 5 wt % FBA DIW Surface tension: 46.0 mN/m droplet lab tensiometer PFH-water interface; pendant drop Fig. 3.2b
PVA in water PFH drop 10 wt % PVA (13-23k) in DIW Surface tension: 37.1 mN/m droplet lab tensiometer PFH-water interface; pendant drop Fig. 3.2b
Combined FBA-PVA condition PFH + 5 wt % FBA 10 wt % PVA (13-23k) in DIW Surface tension: 38.5 mN/m droplet lab tensiometer PFH-water interface; pendant drop Fig. 3.2b
Interfacial mechanics follow-up HFE-7500 with different fluorinated surfactant systems, including HFE-7500 + 5 wt % FBA Water interfaces with 10 wt % PVA and controls Storage modulus, loss modulus, complex shear modulus, complex interfacial viscosity HR 20 discovery hybrid rheometer by TA instruments with DWR setup Amplitude sweep: 0.1% to 10% strain at 10.0 rad/s; frequency sweep: 1.0 to 100 rad/s at 1.0% strain in the LVE region Fig. 3.3; controls include 008-FluoroSurfactant from RAN Biotechnologies, PVA alone, or none
Reconfigurable capsule demonstration 1:1 hexane:PFH shell with FBA polymer at 3–5 wt % Inner and outer phases: 10 wt % PVA aqueous solutions Bright-field and fluorescent confocal capsule morphologies at room temperature and 0 ºC Glass capillary microfluidic device; confocal microscopy Double emulsions collected onto a temperature control stage; phase separation captured during the high-to-low temperature transition Fig. 3.4; hexane phase dyed with 1 mg/ml Nile Red

Key Findings

Interfacial tension dropped with both individual additives

The PFH-water baseline in Figure 3.2b was 71.6 mN/m. Adding 5 wt % FBA to PFH lowered the measured value to 46.0 mN/m, and adding 10 wt % PVA to the aqueous phase lowered it to 37.1 mN/m.

The combined FBA-PVA interface behaved differently from tension reduction alone

The combined FBA-in-PFH/PVA-in-water condition showed a similarly low value of 38.5 mN/m. The study states that lasting stabilization against coalescence was obtained only in the combined FBA-PVA system, which the author attributes to the synergistic interfacial assembly of FBA and PVA.

Elastic interfacial film formation was directly visualized

Transient wrinkle formation during retraction of an HFE-7500 drop containing 3 wt % FBA from a water bath containing 0.08 M HEPES and 2 wt % PVA indicated formation of a solid elastic interfacial film. The slow disappearance of the wrinkles and return of the drop to its original shape supported the dynamic nature of the boronic ester bonds.

Interfacial mechanics increased strongly with the FBA-PVA pair

Interfacial rheology showed that the FBA-PVA interface exhibited a complex shear modulus two orders of magnitude higher than interfaces with 008-FluoroSurfactant from RAN Biotechnologies, PVA alone, or no surfactants. The frequency sweep also showed marked shear-thinning behavior.

The stabilized system supported thermal morphology reconfiguration

After interfacial crosslinking, double emulsion droplets with a water core and a 1:1 hexane:PFH shell became elastic capsules. Because hexane and PFH exhibit temperature-dependent miscibility with an LCST around 23 ºC, cooling from room temperature to 0 ºC led to phase separation and lower-symmetry two-core triple emulsion drops.

Figures & Visuals

What it shows

It shows pendant-drop images and surface-tension values for PFH-water interfaces with no additive, with FBA in PFH, with PVA in water, and with the combined FBA-PVA condition.

What it shows

It shows the boronic-acid coupling chemistry and the transient wrinkling observed when an FBA-containing fluorinated-oil drop retracts from PVA-containing water.

What it shows

It shows DWR interfacial rheology, including the higher modulus and shear-thinning response of the FBA-PVA interface.

What it shows

It shows microfluidic fabrication of the capsules and the morphology change between room temperature and around 0 ºC for the hexane-perfluorohexane system.

Why It Matters

In this study, Dropometer interfacial tension measurements were used to compare interface formulations for hydrocarbon-fluorocarbon-water emulsions. FBA alone, PVA alone, and the combined condition all lowered the PFH-water interfacial tension to different extents, while the thesis reports lasting stabilization against coalescence only for the combined condition.

The rest of the chapter characterizes the combined FBA-PVA interface further, linking it to elastic-film formation, higher interfacial modulus, and temperature-driven reconfiguration of the resulting capsules into more complex morphologies.

Practical Takeaways

Use pendant-drop data comparatively

The surface-tension measurements were used to compare additive placement across the PFH and aqueous phases using the same interface readout.

Low tension was not the full design criterion

In this workflow, several conditions lowered the PFH-water interfacial tension, but durable stabilization tracked with the combined FBA-PVA assembly.

Pair tension with interfacial mechanics

The study combined Dropometer measurements with wrinkle observation and DWR rheology to distinguish elastic-film formation from simple adsorption effects.

Connect interface screening to downstream capsule behavior

The interfacial formulation that stabilized the PFH-water interface was then used in 1:1 hexane:PFH shell capsules that reconfigured on cooling.

FAQ

Frequently Asked Questions

What did the Droplet Lab tensiometer measure in this chapter?

The dissertation states that interfacial tensions were characterized by the pendant drop method on the Droplet Lab tensiometer. The measured interfaces were between perfluorohexane (PFH) and water, with and without the fluorophilic boronic acid (FBA) copolymer in the PFH phase and poly(vinyl alcohol) (PVA) in the water phase (Figure 3.2).

What interfacial tension values were reported?

Without additives the PFH and water interface measured about 72 mN/m. FBA in the PFH phase lowered it to 46 mN/m and PVA in the water phase lowered it to 37 mN/m, while combining both gave a similar value.

What do these results mean for emulsion stability?

The author reports that lasting stabilization against coalescence was obtained only with the combined FBA and PVA system, even though each additive alone also lowered the interfacial tension. The chapter attributes stabilization to a synergistic interfacial assembly, supported by wrinkle observations and double wall ring interfacial rheology rather than by tension values alone. Pendant drop values for the hexane and PFH interface used in the capsule shells are not reported.