Not all contact-angle measurements are created equal: the right method depends on the application. A sessile drop and a tilted plate can give very different, and equally correct, numbers.
AATCC TM118 grades how well a finished fabric resists wetting by eight hydrocarbon liquids of falling surface tension, judged by eye. The Dropometer is partially compliant: it measures the contact angle of each test oil on the fabric with the fabric holder, and checks each liquid's surface tension by pendant drop with ±0.03 mN/m accuracy (Dropometer datasheet). Contact angle is an alternative method, not the visual grading method the standard specifies; if a contract or customer requires an AATCC TM118 result, run a correlation study on your own samples first.
What brings you here?
Quality manager
You grade fabric to AATCC TM118 by eye. Adding the angle of each test oil gives you a number you can trend by lot and by wash count.
R&D chemist
You need to compare finishes, including fluorine free ones, that land on the same grade. The angle of each test oil shows which one holds up better.
Lab technician
A grade is only as good as the liquid behind it. Check each hydrocarbon by pendant drop before it touches a fabric, especially the volatile octane and heptane.
Free one page checklist for AATCC TM118 with the Dropometer: how to check the hydrocarbon test liquids, the settings to lock, how to read oil contact angles next to the visual grade and what to record. Made for textile QC and finish labs.
Talk to an expert
Tell us your fabric, finish and the oil grade your customer asks for. We will check a test liquid and measure oil contact angles on your fabric live, and tell you plainly where the Dropometer fits next to the visual grade and where it does not.
AATCC TM118-2020e, Test Method for Oil Repellency: Hydrocarbon Resistance Test, from AATCC, the American Association of Textile Chemists and Colorists. It detects a fluorochemical or other low energy finish by rating how well the fabric resists wetting by hydrocarbons of different surface tensions. AATCC lists it as technically equivalent to ISO 14419.
Partially compliant. The Dropometer measures the wetting the standard is after, as the contact angle of each test oil on the fabric, and checks each liquid's surface tension by pendant drop. Contact angle is an alternative method, not the visual grading method the standard specifies. If a contract or customer requires an AATCC TM118 result, run a correlation study on your own samples first.
Contact angle range 10° to 175°, accuracy ±0.35°, display resolution 0.01°. Pendant drop surface tension up to 75 mN/m, accuracy ±0.03 mN/m. Both from the Dropometer product specification. The test oils, about 20 to 31.5 mN/m, sit inside that range.
The drop shape method is validated against a KRUSS DSA100E in two peer reviewed papers. See our validation publications.
Researchers used a Droplet Lab tabletop goniometer to measure water contact angle on electrospun polystyrene fibre mats, with values from 130.1° to 143.2° (Macromolecular Rapid Communications, 2024). Read our analysis of the paper. We have no published client study with hydrocarbon liquids or run to AATCC TM118 itself.
The grade and the angle show resistance to wetting by the test oils. AATCC notes the grade is not an absolute measure of resistance to staining by all oily materials, and ISO 14419 does not cover penetration by oil based chemicals. Pair them with the test your product needs.
What this standard measures, who uses it and how the Dropometer fits
AATCC TM118 is the oil repellency test for fabric: drops of hydrocarbon liquids, each with a lower surface tension than the last, go on the fabric, and the grade is the highest numbered liquid that does not wet it. Mills, finish makers and brands of workwear, upholstery and protective clothing use it to check oil and stain repellent finishes, often after washing or dry cleaning. The Dropometer is partially compliant. It checks the surface tension of each test liquid by pendant drop and measures the contact angle of each liquid on the fabric, but it does not give the visual grade. Contact angle is an alternative method, not the visual grading method the standard specifies; if a contract or customer requires an AATCC TM118 result, run a correlation study on your own samples first.
The eight liquids run from white mineral oil at 31.5 mN/m down to n-heptane at about 20 mN/m (AATCC 118-1997, Note 12.2; surface-tension.de). All of them sit well inside the Dropometer's 75 mN/m pendant drop range. A drop that stays clear and rounded for 30 s passes. A drop that darkens the fabric or wicks in fails. Oils partly wet most flat surfaces, and the Wenzel model shows that roughness alone makes such a liquid wet even more. A fabric holds an oil drop only when its fibres trap air under the drop, so the angle you read is an apparent angle.
The Dropometer contact angle and surface energy kit reads each test oil on the fabric holder with ±0.35° accuracy, and its pendant drop mode checks the liquids before you test. For the wider fabric workflow, see textile and DWR water repellency testing. Read our federal research lab coated fabric case study, and our analysis of a study where researchers measured water contact angle on electrospun polystyrene fibre mats on a Droplet Lab goniometer. For background, read the fabrics guide and the contact angle measurement guide.
Further reading: reproducible contact angle measurement, choosing a contact angle goniometer for industrial QA, a classroom experiment on superhydrophobic glass and reference values for water contact angle on common surfaces.
Check the AATCC TM118 test oils, then measure each one on the fabric, step by step
These steps describe a pendant drop check and a contact angle measurement on the Dropometer. They do not replace the visual grading in AATCC TM118. Buy the current edition and follow its liquids, drop size, timing and grading rules for any result you report to the standard.
From the Dropometer product specification. Display resolution is 0.01°.
Pendant drop surface tension accuracy, from the Dropometer product specification. Range up to 75 mN/m.
Base plate, phone holder, LED, syringe and fabric holder. No PC needed.
Use the hydrocarbon liquids the standard lists, from white mineral oil to n-heptane, and mix liquid 2 by volume. Cap and label each bottle with liquid number and date. Note each liquid's density at room temperature from the supplier's data, because the pendant drop fit needs it.
Hang a drop of each liquid from a clean needle and read its surface tension. Compare it with the value the standard lists and record room temperature. These oils give small, light drops, so use a needle that gives a well formed drop, and read octane and heptane quickly before they evaporate.
Condition the specimens as the standard specifies. Clamp each one flat and face up in the fabric holder, without stretching it. Test thin, open fabrics in two layers, as the 1997 edition advises. Use silicone free gloves, because silicone changes the result.
Start with liquid 1 and work up, as in the standard. Place the drop size the standard specifies, keep it the same for every drop and do not touch the fabric with the needle. Use a fresh spot for each drop and rinse the syringe between liquids.
Capture each drop at the observation time the standard sets, 30 s in the 1997 edition. Automatic edge and baseline detection reads the angle, so two operators get the same reading from the same image. Also note what you see by eye: a clear rounded drop, partial darkening or wicking. The Cassie Baxter model explains why air under the drop keeps the angle high.
Record the visual grade to the standard, the mean angle and spread for each liquid, wash count, room temperature and the liquid check values. State that the angles are contact angle results, not the AATCC TM118 grade. Use your correlation study to link the two.
This video shows a complete Dropometer contact angle measurement, from setup to result. It covers the steps above in one run.
Independent benchmarking and publication-based validation references.
Peer reviewed method
The Dropometer drop shape method is validated against a KRUSS DSA100E in two peer reviewed papers, one in Review of Scientific Instruments and one in Colloids and Surfaces A. Published accuracy is ±0.35° for contact angle and ±0.03 mN/m for surface tension (Dropometer product specification).
Customer evidence
A university team used a Droplet Lab tabletop goniometer to measure water contact angle on electrospun polystyrene fibre mats, published in Macromolecular Rapid Communications in 2024. We have no published client study with hydrocarbon liquids or run to AATCC TM118 itself.
AATCC TM118 · oil repellency · textile finish QC
Each grade depends on a test liquid of the right surface tension. The public 1997 edition lists, at 25 °C, 31.5 mN/m for liquid 1 (Kaydol white mineral oil), 27.3 for n-hexadecane (3), 26.4 for n-tetradecane (4), 24.7 for n-dodecane (5), 23.5 for n-decane (6) and 21.4 for n-octane (7) (source: AATCC 118-1997, Note 12.2). The value for n-heptane (8) is not legible in the public scan; n-heptane reads 20.14 mN/m at 20 °C (source: surface-tension.de). The edition gives no value for liquid 2, the 65:35 Kaydol and n-hexadecane mix, so set your own reference from a fresh batch.
These liquids lose about 0.09 mN/m for every °C warmer (same source), so check them by pendant drop at a recorded room temperature and compare like with like. Liquids 7 and 8 evaporate fast: keep bottles capped and read the drop promptly.
AATCC TM118 gives a grade, not a contact angle limit. Your specification or your customer sets the minimum grade. To use angles, run a correlation study: grade a set of fabrics by eye to the standard, measure the angle of the same liquids on the same fabrics, then set your angle limits from that data. See reproducible contact angle measurement for bench practice.
Example data for illustration only. These are not measured results. Your own fabric, finish and wash procedure decide the numbers.
| Liquid | Reference at 25 °C (mN/m) | Pendant drop check at 21 °C (mN/m) | Angle as finished (°) | Angle after 10 washes (°) | Visual result after 10 washes |
|---|---|---|---|---|---|
| 3: n-hexadecane | 27.3 | 27.63 | 126.4 | 112.3 | Pass: clear, rounded drop |
| 4: n-tetradecane | 26.4 | 26.75 | 122.8 | 103.5 | Pass: clear, rounded drop |
| 5: n-dodecane | 24.7 | 25.04 | 116.5 | 86.1 | Borderline: partial darkening |
| 6: n-decane | 23.5 | 23.87 | 108.9 | Wicked in, no angle | Fail: wetting |
| 7: n-octane | 21.4 | 21.78 | Wicked in, no angle | Wicked in, no angle | Fail: wetting |
Each angle is the mean of five drops read at 30 s on the fabric holder. The liquids were checked by pendant drop at 21 °C room temperature, so they read about 0.35 mN/m above the 25 °C reference values (AATCC 118-1997; temperature coefficients from surface-tension.de). Visual grade: 6 as finished and 4.5 after 10 washes, against an example customer minimum of 5. The falling angles of liquids 3 and 4 show the finish wearing before the grade drops.
The same example data as the table: oil contact angles fall after washing, and liquid 6 wicks in, so the visual grade drops from 6 to 4.5.
Example data, not measured results.
What changes when a measured oil contact angle sits next to the visual grade
| Metric | Before Dropometer | With Dropometer | Results |
|---|---|---|---|
| How wetting is judged | Drop shape read by eye, one grade per fabric | Angle of each test oil read from the drop image | No published benchmark |
| Test liquid check | Bought or mixed and trusted | Surface tension of each oil read by pendant drop, ±0.03 mN/m accuracy | Dropometer product specification |
| Change after washing | Seen only when the grade drops | A falling angle within the same grade | No published benchmark |
| Reported AATCC TM118 grade | Visual grading to the standard | Still needs visual grading to the standard | AATCC TM118 scope |
Bring a fabric as finished and the same fabric after washing. We will measure both live with your hydrocarbon test liquids.
Enter your own numbers. The defaults are examples, not Droplet Lab results.
Quality cost snapshot
Result
Annual exposure combines audit/CAPA cost, customer rejections, and recall probability; avoided exposure applies the risk reduction percentage.
Lock these before you compare angles between lots, washes or sites
| Parameter | Recommended Setting | Technical Rationale |
|---|---|---|
| Method | Sessile drop contact angle of each test oil, reported next to the visual grade | AATCC TM118 specifies a visual grade. The angle is an alternative reading. |
| Test liquids | The eight hydrocarbons the standard lists, liquid 2 mixed by volume | Each grade depends on the liquid's surface tension. |
| Liquid check | Pendant drop, density entered for each liquid at room temperature | The Young-Laplace fit needs the density of the drop. All liquids sit inside the 75 mN/m range. |
| Specimen mounting | Fabric holder, flat and face up, not stretched; two layers for thin, open fabric | A wavy or see through fabric changes the baseline and the angle. |
| Drop size | As the standard specifies, fixed for every drop | A changing size adds spread that is not about the finish. Drops come from the needle, not a dropping bottle: note it. |
| Observation time | As the standard specifies, fixed for every drop | Low surface tension oils can wick in over time. The 1997 edition uses 30 s. |
| Fit model | Young-Laplace or polynomial, the same for every run | Both models are in the Dropometer software. Do not mix them within one study. |
| Replicates | Several drops per liquid, each on a fresh spot | Weave and finish vary across the fabric. The 1997 edition grades on five drops. |
| Environment | Record room temperature; condition samples and liquids first | The Dropometer has no temperature control and runs at 10 to 45 °C. |
Pick the method by the question you need answered
Start: What do you need to know about the finished fabric?
Your customer or specification asks for a minimum oil repellency grade.
Grade the fabric by eye as the standard specifies. Contact angle on the Dropometer is an alternative method: run a correlation study on your own samples before you use angles against a TM118 grade.
You mix liquid 2 yourself, or a bottle of octane or heptane has been open for a while.
Check each liquid by pendant drop on the Dropometer and compare it with the value the standard lists, at a recorded room temperature.
You want to compare finishes with the same grade, or see oil repellency fade with washing before the grade drops.
Measure the contact angle of each test oil on the fabric holder. See textile and DWR water repellency testing.
You need resistance to water and alcohol stains, to sprayed water, or how fast a fabric takes up water.
Use AATCC TM193 for water and alcohol, AATCC TM22 for the spray test, or AATCC TM 79 for absorbency. For water resistance under hydrostatic pressure, use AATCC TM127.
Start from the question you need answered. Blue boxes are measurements the Dropometer runs.
Is AATCC TM118 the right test?
What the oil angle tells you next to the grade, and what it does not
Honest limits of the method and of the Dropometer
It rates how well a finished fabric resists wetting by hydrocarbon liquids of falling surface tension, from white mineral oil to n-heptane. Drops of each liquid go on the fabric in turn. The oil repellency grade is the highest numbered liquid that does not wet the fabric, from 0 to 8.
In part, so it is partially compliant. It checks each test liquid's surface tension by pendant drop and measures each oil's contact angle on the fabric with the fabric holder. It does not give the visual grade. Contact angle is an alternative method, not the visual grading method the standard specifies; if a contract or customer requires an AATCC TM118 result, run a correlation study on your own samples first.
Yes. The liquids span about 20 to 31.5 mN/m (source: AATCC 118-1997; surface-tension.de), and pendant drop reads up to 75 mN/m with ±0.03 mN/m accuracy (Dropometer datasheet). On a repellent finish, passing drops sit far above the 10° lower limit for contact angle. A drop that wicks in is recorded as wetting.
AATCC lists TM118 as technically equivalent to ISO 14419. ISO published the third edition, ISO 14419:2025, in October 2025 and has already marked it for revision. Check which designation and edition your customer names, and quote that one in your test report.
Oils have a low surface tension, so only a very low energy surface keeps them beaded. A 2019 study from the University of Leeds and Stockholm University found fluorine free finishes gave good rain protection but no repellency to oil based stains (Schellenberger et al., Journal of Cleaner Production). Oil angles on the Dropometer help compare such finishes in development.
Oils partly wet most flat surfaces, and the Wenzel model shows roughness alone makes that worse. Fabric holds the drop when fibres trap air beneath it, as the Cassie Baxter model describes. Compare angles only between fabrics of the same construction.
No. The Dropometer has no temperature control and runs at 10 to 45 °C. Each test oil loses about 0.09 mN/m per °C warmer (source: surface-tension.de), so condition samples and liquids as the standard specifies and record room temperature with every result.
This page summarises AATCC TM118 in our own words and does not reproduce the standard. Buy the current edition from AATCC before you write a procedure. The Dropometer is partially compliant: it checks the hydrocarbon test liquids by pendant drop and measures their contact angle on the fabric, an alternative to the visual grading the standard specifies. Droplet Lab does not certify compliance. Your own method validation does.
Sources, review and tools
AATCC store: TM118 Test Method for Oil Repellency: Hydrocarbon Resistance Test (technically equivalent to ISO 14419); Thai DSS library record: AATCC TM118-2020e (2022); AATCC Test Method 118-1997 (public copy): liquids, surface tensions, 30 s observation, grading; ISO 14419:2025 listing, the Dropometer product specification and Droplet Lab validation papers.
Drafted with Claude (Anthropic) from the sources listed, then reviewed by Droplet Lab.
We did not copy text from the standard. Buy the current edition for the full procedure.